Becton Dickinson recalls BD Max CT/GC/TV kits with foil bags that may not seal properly. This could lead to contamination of reagents. Affected kits have specific lot numbers and expiration dates.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
Becton Dickinson is recalling BD Max CT/GC/TV kits because the foil bags containing extraction tubes might not seal properly. This could cause contamination of the reagents, potentially leading to inaccurate test results.
The foil bags may not seal properly, which could allow contaminants to enter the extraction tubes. This might lead to inaccurate test results if the reagents are used in diagnostic testing.
Users of BD Max CT/GC/TV kits who purchased the recalled lots between 2019 and 2020 are affected. Laboratory personnel and healthcare providers using these kits are most at risk.
Check the lot numbers on the kit packaging: 9080876, 9092733, 9092734, or 9106875. The kits expire by April 14, 2020, or May 12, 2020.
Look for the lot numbers 9080876, 9092733, 9092734, or 9106875 on the packaging.
The foil bags containing extraction tubes may not seal properly, which could lead to contamination of the reagents.
Check the lot numbers on the packaging: 9080876, 9092733, 9092734, or 9106875. The kits expire by April 14, 2020, or May 12, 2020.
The recall does not specify a stop-use action, so you can continue using them if you have checked the lot numbers.
We’ve drafted a message you can send Becton Dickinson to request your refund or repair — edit it as you like.
Subject: Recall Z-2331-2019 — Becton Dickinson Becton Dickinson Hello, I own a Becton Dickinson that is covered by recall Z-2331-2019 from Becton Dickinson. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.