Becton Dickinson recalls BD Max StaphSR Kit (Catalog #443418) due to foil bags for extraction tubes possibly not sealing properly. Affected units include specific lots and expiration dates.
The foil bags containing extraction tubes for BD MAX" Reagent products may not been sealed properly.
Becton Dickinson is recalling BD Max StaphSR kits with catalog number 443418 because the foil bags containing extraction tubes might not seal properly. This could lead to contamination or improper use of the reagents. If you own this kit, stop using it immediately and contact Becton Dickinson for a replacement.
The foil bags that hold extraction tubes for BD MAX reagent products may not seal properly. This could allow contaminants to enter the tubes or cause improper handling of the reagents, potentially leading to inaccurate test results or contamination.
Users of BD Max StaphSR kits with catalog number 443418 and specific lot numbers (9086668 and 9092724) that expired by September 2, 2020, are affected. Laboratory staff and healthcare professionals using these kits are most at risk.
Check for the catalog number 443418 on the kit. Look for lot numbers 9086668 or 9092724 with an expiration date of September 2, 2020.
Immediately stop using the BD Max StaphSR kit if it has the catalog number 443418 and lot numbers 9086668 or 9092724.
The catalog number is 443418.
Lot numbers 9086668 and 9092724 are affected.
The expiration date is September 2, 2020.
We’ve drafted a message you can send Becton Dickinson to request your refund or repair — edit it as you like.
Subject: Recall Z-2332-2019 — Becton Dickinson Becton Dickinson Hello, I own a Becton Dickinson that is covered by recall Z-2332-2019 from Becton Dickinson. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.