Zimmer Biomet recalls Vanguard XP Tibial Tray 79mm (Item #195252) due to locking bar not fully engaging. Affected items have specific lot numbers. Consumers should contact Zimmer Biomet for details.
The locking bar not fully engaging
Zimmer Biomet is recalling Vanguard XP Tibial: 79mm Tibial Trays with Item #195252 that may not lock securely. This could lead to improper joint alignment during surgery, risking complications for patients.
The locking bar may not fully engage, which could cause the implant to shift during surgery. This might lead to improper joint alignment and potential complications for the patient.
Orthopedic surgeons and their patients who received the recalled Vanguard XP Tibial Tray 79mm (Item #195252) during surgery. Patients are most at risk if the implant does not lock properly during procedures.
Check for the product name 'Vanguard XP Tibial Tray 79mm' and Item #195252. Affected items have specific lot numbers: 332750, 833230, 515120, 752550, 515130, 373090, 919280, 754510, 764430, 764440, 861230, 837810, 846560, 907730, 928810, 907720, 907740, 969000, 969010, 590620, 590600, 663320, 663330, 663340, 590610, 571010, 663310.
Reach out to Zimmer Biomet directly for specific guidance on the recalled item and next steps.
The recalled item is Vanguard XP Tibial Tray 79mm with Item #195252.
Check for the product name 'Vanguard XP Tibial Tray 79mm' and Item #195252. Affected items have specific lot numbers listed in the recall notice.
Contact Zimmer Biomet directly for details on the recall and next steps.
We’ve drafted a message you can send Zimmer Biomet to request your refund or repair — edit it as you like.
Subject: Recall Z-2535-2019 — Zimmer Biomet Zimmer Biomet Hello, I own a Zimmer Biomet that is covered by recall Z-2535-2019 from Zimmer Biomet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.