Quality Electrodynamics recalls Siemens 10606524 knee coils due to potential ferrous material in pushbutton assembly. Affected units may have safety risks during MRI use.
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
Quality Electrodynamics is recalling Siemens 10606524 knee coils because small metal pieces could be inside the pushbutton. This poses a risk during MRI scans, as metal can cause burns or other injuries.
The pushbutton assembly may contain small metal pieces that could become dislodged during MRI scans. This metal could heat up and cause burns or other injuries to the patient.
MRI patients using these knee coils are at risk. The recall affects specific models sold with serial numbers starting with 1433 1702 1440.
Check for the Siemens model number 10606524 on the coil. Affected units have serial numbers starting with 1433 1702 1440.
Look for the Siemens model number 10606524 on the coil.
Contact Quality Electrodynamics for a replacement or repair if you have the affected model.
Yes, the pushbutton may contain small metal pieces that could cause burns during MRI scans.
Check for the Siemens model number 10606524 and serial numbers starting with 1433 1702 1440.
No, but you should contact Quality Electrodynamics if you have the affected model to get a replacement.
We’ve drafted a message you can send Quality Electrodynamics to request your refund or repair — edit it as you like.
Subject: Recall Z-0034-2020 — Quality Electrodynamics Quality Electrodynamics Hello, I own a Quality Electrodynamics that is covered by recall Z-0034-2020 from Quality Electrodynamics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.