Quality Electrodynamics recalls Siemens 10606525 knee coils with possible ferrous material in pushbutton assembly. Affected units have specific serial numbers. No remedy stated in recall notice.
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
Quality Electrodynamics recalls certain Siemens 3T knee coils due to potential ferrous material in the pushbutton assembly. This could cause interference with medical devices if the coil is used near MRI machines.
The pushbutton assembly might contain unintended ferrous material, which could interfere with MRI machines. This could lead to safety issues during medical imaging procedures if the coil is used improperly.
Patients using MRI-compatible knee coils for medical imaging may be affected. Those with specific serial numbers (1571-1575) are most at risk.
Check for the Siemens model number 10606525 and serial numbers 1571 through 1575 on the product. The product is sold by Quality Electrodynamics.
Verify the Siemens model number 10606525 and serial numbers 1571-1575 on the product.
The recall notice does not state any remedy or fix for this issue.
The recall notice does not specify any action steps for affected products.
Check for the Siemens model number 10606525 and serial numbers 1571 through 1575 on the product.
We’ve drafted a message you can send Quality Electrodynamics to request your refund or repair — edit it as you like.
Subject: Recall Z-0035-2020 — Quality Electrodynamics Quality Electrodynamics Hello, I own a Quality Electrodynamics that is covered by recall Z-0035-2020 from Quality Electrodynamics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.