Centurion recalls disposable circumcision clamps due to potential foreskin removal without scalpel use. Affected units sold in 2019 with batch 2019061201.
A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.
Centurion recalls disposable circumcision clamps that may remove foreskin without a scalpel. This could cause injury during infant circumcision procedures.
The clamp's hole angle doesn't match the bell's angle, which may cause the foreskin to be removed without a scalpel. This could lead to injury during circumcision procedures.
Parents of infants undergoing circumcision are most at risk. The product is used by medical professionals during infant circumcision procedures.
Check for the model 'CIRCLAMP W/1.3 CM BELL' with batch number 2019061201. The product was sold in 2019 for infant circumcision procedures.
Look for batch number 2019061201 on the clamp packaging or label.
The recall indicates a potential risk of foreskin removal without a scalpel, so it is not safe to use.
Check for the model 'CIRCLAMP W/1.3 CM BELL' with batch number 2019061201.
The recall does not specify a remedy, so contact the manufacturer for guidance.
We’ve drafted a message you can send Centurion to request your refund or repair — edit it as you like.
Subject: Recall Z-0121-2020 — Centurion Centurion Hello, I own a Centurion that is covered by recall Z-0121-2020 from Centurion. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.