Alto Development Corp recalls Thorecon Plating System Kits Model 94-1400-04 for sternum fracture repair due to defective tensioners that may not function properly.
Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable.
Alto Development Corp is recalling Thorecon Plating System Kits Model 94-1400-04 used for sternum fracture repair. The kits have a defect that makes the tensioners inoperable, which could lead to improper healing or complications.
The tensioners in the kits have a molding defect that blocks the cable channel, making them unable to function properly. This could result in improper bone fixation and potential complications during healing.
Patients undergoing sternum fracture repair using the recalled Thorecon Plating System Kits Model 94-1400-04 are affected. Surgeons and medical staff using these kits for procedures are at risk.
Check for the model number 94-1400-04 on the kit packaging or label. The kits are used for sternum fracture repair and were sold with specific serial numbers listed in the recall.
Verify the model number on the kit packaging or label to confirm if it matches the recalled model 94-1400-04.
The Thorecon disposable tensioners in these kits have a molding defect that blocks the cable channel, making them inoperable.
The recall record does not specify a remedy, so contact Alto Development Corp directly for guidance.
Yes, the recall includes specific serial numbers: 327705, 327706, 330587, 330588, 330753, 334396, 345417, 345418.
We’ve drafted a message you can send Alto Development Corp to request your refund or repair — edit it as you like.
Subject: Recall Z-0244-2020 — Alto Development Corp Alto Development Corp Hello, I own a Alto Development Corp that is covered by recall Z-0244-2020 from Alto Development Corp. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.