Theralase recalls TLC-2:000 Therapeutic Medical Laser Systems with serial numbers 3333-3350 due to electromagnetic interference exceeding safety standards. Affected units may pose risks to medical equipment.
Laser Probe (Model Number: TLC-2001) Electromagnetic Interference ( EMI ) in excess of IEC-60601-1-2 (4th edition) standard
Theralase is recalling TLC-2000 Therapeutic Medical Laser Systems with serial numbers 3333 to 3350 because they may cause electromagnetic interference that could affect other medical devices. This could lead to potential safety issues during medical treatments.
The laser probe (model TLC-2001) in the system may emit electromagnetic interference that exceeds the IEC-60601-1-2 (4th edition) safety standard. This interference could potentially disrupt other medical devices during use, though the risk is not severe.
Medical professionals using the TLC-2000 system with serial numbers 3333 to 3350 are most at risk. Patients receiving treatments with this equipment may also be affected.
Check if your TLC-2000 system has a serial number between 3333 and 3350 inclusive. The laser probe model number is TLC-2001.
Verify your TLC-2000 system's serial number to see if it falls between 3333 and 3350.
The recall is for electromagnetic interference that may affect other medical devices, but the risk is low and not severe.
Check your serial number; if it's between 3333 and 3350, contact Theralase for further instructions.
The affected laser probe model number is TLC-2001. Check your product's model number.
We’ve drafted a message you can send Theralase to request your refund or repair — edit it as you like.
Subject: Recall Z-1266-2020 — Theralase Theralase Hello, I own a Theralase that is covered by recall Z-1266-2020 from Theralase. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.