Energetiq recalls 15 EQ-400-RH-QZ and EQ-400-LH-BK broadband light products due to unsafe laser collimator removal risk. Affected units made between June 2016 and September 2019.
On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.
Energetiq is recalling 15 specific models of broadband light products because a safety feature that shuts down the system when the laser collimator is removed doesn't work reliably. This could expose people to unsafe light exposure if the device is operated without proper safety checks.
The device has a safety interlock that fails to shut down the system when the laser collimator is removed during operation. This could allow unsafe light exposure to people nearby or working on the device.
Workers using these lights for industrial or scientific purposes are most at risk. The recall affects 15 units manufactured between June 2016 and September 2019.
Check for the model numbers EQ-400-RH-QZ or EQ-400-LH-BK on the product. These units were made between June 3, 2016 and September 26, 2019.
Look for the model numbers EQ-400-RH-QZ or EQ-400-LH-BK on the device.
Energetiq Technology Inc will provide guidance on next steps for affected units.
The safety interlock that shuts down the system when the laser collimator is removed doesn't function reliably, creating a potential safety hazard.
15 units manufactured between June 3, 2016 and September 26, 2019.
Check the model number and contact Energetiq Technology Inc for guidance on next steps.
We’ve drafted a message you can send Energetiq to request your refund or repair — edit it as you like.
Subject: Recall Z-0870-2020 — Energetiq Energetiq Hello, I own a Energetiq that is covered by recall Z-0870-2020 from Energetiq. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.