Cook recalls Torcon NB Advantage Catheters due to potential sterility compromise from undersealed packaging. Affected units have specific lot numbers and UDI codes. Consumers should check their product details to determine if recalled.
The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
Cook is recalling Torcon NB Advantage Catheters because the packaging may not seal properly, risking contamination. This could lead to infections or other complications if the catheters are used without proper sterility.
The catheters' packaging pouches may have an undersealed chevron, which could compromise sterility during use. This risk is specific to the packaging and not the catheter itself.
Physicians using Torcon NB Advantage Catheters for vascular procedures are most at risk. The recall affects specific lot numbers and UDI codes of the catheters.
Check for lot number NS9708132 or UDI code (01)00827002059252(17)220513(10)NS9708132 on the packaging. The catheters are intended for use in vascular procedures by trained physicians.
Look for lot number NS9708132 or UDI code (01)00827002059252(17)220513(10)NS9708132 on the packaging.
The packaging pouches may have an undersealed chevron, which could compromise sterility during use.
Check the lot number or UDI code on the packaging to see if it's affected. If it is, contact Cook for further instructions.
There is a potential risk of contamination if the packaging is compromised, which could lead to infections or complications.
We’ve drafted a message you can send Cook to request your refund or repair — edit it as you like.
Subject: Recall Z-0430-2020 — Cook Cook Hello, I own a Cook that is covered by recall Z-0430-2020 from Cook. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.