Integra recalls C7000 Clarity Ultrasonic Tissue Ablation System's 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) due to insufficient drying time in 30-minute cycles, risking incomplete sterilization.
The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operators manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.
Integra is recalling a specific sterilization tray for their C7000 Clarity system because it doesn't dry properly in 30-minute cycles as required. This could lead to incomplete sterilization and potential infection risks if used without proper drying.
The sterilization tray requires 40 minutes for complete drying but the system's manual states 30 minutes. This mismatch means the tray may not dry fully, increasing the risk of incomplete sterilization and potential infection.
Medical professionals using the C7000 Clarity system with the specific sterilization tray model (C7136 Rev C / 72904491 Rev A) are affected. Those in settings where the tray is used for surgical procedures face the highest risk.
Check for the model number C7136 Rev C or 72904491 Rev A on the sterilization tray. The product was sold with the Croatian OPS Manual and part number 60905214.
Identify the model number on the tray (C7136 Rev C or 72904491 Rev A) to confirm if it's affected.
The tray does not meet the required drying time for 30-minute cycles as stated in the manual, which could lead to incomplete sterilization.
The recall record does not specify a remedy, so contact Integra LifeSciences for guidance.
The hazard involves potential incomplete sterilization, but the recall indicates a low risk as the issue is specific to drying time and not immediate safety risks.
We’ve drafted a message you can send Integra LifeSciences to request your refund or repair — edit it as you like.
Subject: Recall Z-0650-2020 — Integra LifeSciences Integra LifeSciences Hello, I own a Integra LifeSciences that is covered by recall Z-0650-2020 from Integra LifeSciences. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.