Cook recalls Torcon NB Advantage Catheters due to potential sterility compromise from undersealed packaging. Affected units may have contaminated products that could cause infection during medical procedures.
The potential compromise of the sterility of products from an undersealed chevron in the packaging pouches of Torcon NB Advantage Catheters.
Cook is recalling Torcon NB Advantage Catheters because their packaging may have an undersealed chevron, risking contamination. This could lead to infections during medical procedures if the catheters are used.
The packaging pouches may have an undersealed chevron, which could compromise the sterility of the catheters. If the catheters are used without proper sterility, they might introduce infections during medical procedures.
Physicians using these catheters for vascular procedures are most at risk. The recall affects specific lot numbers and model types used in angiographic procedures.
Check for lot number NS9717356 or UDI (01)00827002471689(17)220516(10)NS9717356 on the packaging. These catheters are used in vascular procedures by trained physicians.
Look for lot number NS9717356 or UDI (01)00827002471689(17)220516(10)NS9717356 on the packaging.
The packaging pouches may have an undersealed chevron, which could compromise the sterility of the catheters.
Check for lot number NS9717356 or UDI (01)00827002471689(17)220516(10)NS9717356 on the packaging. If you have the affected lot, contact Cook for guidance.
The recall is for potential sterility compromise, which could lead to infections during medical procedures, but not a direct risk of serious injury.
We’ve drafted a message you can send Cook to request your refund or repair — edit it as you like.
Subject: Recall Z-0421-2020 — Cook Cook Hello, I own a Cook that is covered by recall Z-0421-2020 from Cook. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.