Polymer Technology Systems recalls refurbished CardioChek Plus analyzers without boxes from 2016-2019 due to battery insertion risks causing overheating. Affected units have specific reference numbers and serial dates.
One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.
Polymer Technology Systems is recalling refurbished CardioChek Plus Professional Analyzers without boxes that were sold between November 2016 and October 2019. The recall is because certain batteries can cause overheating if inserted incorrectly due to a small dimensional difference in the negative button terminal.
The hazard occurs when batteries with a small dimensional difference in the negative button terminal are inserted incorrectly. This can cause the battery terminals to make reverse contact, potentially leading to overheating. However, the risk is low as it only affects specific battery types and requires incorrect insertion.
Users of the refurbished CardioChek Plus analyzers without boxes or carrying cases that were purchased between November 2016 and October 2019 are affected. Most at risk are those who use the device with batteries that have the specific reference numbers mentioned in the recall.
Check for reference numbers 2702 on the analyzer or serial numbers shipped between November 1, 2016, and October 25, 2019. The device should not have a box or carrying case.
Ensure batteries are inserted correctly to avoid reverse contact and overheating.
The recall is due to a small dimensional difference in the negative button terminal of certain batteries that can cause reverse contact when inserted incorrectly, potentially leading to overheating.
Refurbished CardioChek Plus Professional Analyzers without boxes or carrying cases with reference numbers 2702 and serial numbers shipped between November 1, 2016, and October 25, 2019.
No, the recall does not require immediate discontinuation. Users should check battery insertion to prevent overheating.
We’ve drafted a message you can send Polymer Technology Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-0833-2020 — Polymer Technology Systems Polymer Technology Systems Hello, I own a Polymer Technology Systems that is covered by recall Z-0833-2020 from Polymer Technology Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.