Terumo recalls specific roller pumps that may fail due to unusual noise, risking pump stoppage during blood circulation procedures. Affected units have serial numbers starting with 882298.
These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.
Terumo is recalling certain roller pumps used in blood circulation procedures because faulty assembly may cause them to stop working. This could lead to interrupted blood flow during critical medical procedures if the pump fails.
The pumps have a potential failure point where the shaft connects to the locknut. If the shaft end moves freely, it creates unusual noise and can eventually cause the pump to stop working by damaging internal parts. This could interrupt blood flow during critical medical procedures.
Medical professionals using the Terumo Advanced Perfusion Assembly Pump Guts for blood circulation procedures are most at risk. Patients undergoing these procedures could be affected if the pump fails.
Check if your pump has the serial numbers listed: 882298, 882299, 882300, 882752, 882753, 882794, 882795, 882796, 882797, 882798, 882799, 885547, 885548, 885549, 885550, 885553.
Look for the serial numbers listed in the recall notice: 882298 through 885553.
The pumps were assembled with shafts that may fail at the thread where a locknut is installed, causing unusual noise and potential pump stoppage during blood circulation procedures.
Prolonged use in this condition could lead to pump stoppage, resulting in inability to create blood flow due to pulley loosening and drive belt damage.
Check if your pump has a serial number starting with 882298, 882299, 882300, 882752, 882753, 882794, 882795, 882796, 882797, 882798, 882799, 885547, 885548, 885549, 885550, or 885553.
We’ve drafted a message you can send Terumo to request your refund or repair — edit it as you like.
Subject: Recall Z-1110-2020 — Terumo Terumo Hello, I own a Terumo that is covered by recall Z-1110-2020 from Terumo. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.