Arrow recalls FlexTip Plus Epidural Catheterization Sets with potential dural puncture risk. Affected lots include 71F19B2081 through 71F19J1685. Contact Arrow for details.
There is the potential for various failures with the LOR (loss of resistance) syringes included in the kits resulting in dural puncture with some also requiring a blood patch.
Arrow is recalling FlexTip Plus Epidural Catheterization Sets that may cause dural puncture due to LOR syringe failures. This could require a blood patch and poses a risk of serious complications during epidural procedures.
The LOR syringes in the kits can fail, leading to dural puncture. This may cause spinal fluid leakage and sometimes requires a blood patch to seal the leak.
Healthcare professionals using the FlexTip Plus Epidural Catheterization Sets for epidural procedures are most at risk. Patients receiving epidural anesthesia may also be affected if the device fails.
Check the product code on the packaging. Affected lots are 71F19B2081, 71F19C1655, 71F19E2128, 71F19G2444, 71F19B2586, 71F19E1245, 71F19F1432, and 71F19J1685.
Look for the product code on the packaging to identify if it's in the affected lots listed.
The LOR syringes may fail, causing dural puncture which can require a blood patch and lead to serious complications.
Check the product code on the packaging. Affected lots include 71F19B2081 through 71F19J1685.
The recall does not specify a remedy, so contact Arrow directly for guidance.
We’ve drafted a message you can send Arrow to request your refund or repair — edit it as you like.
Subject: Recall Z-1171-2020 — Arrow Arrow Hello, I own a Arrow that is covered by recall Z-1171-2020 from Arrow. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.