RaySearch Laboratories recalls RayStation/RayPlan 9B software build 9.1.0.933 due to potential incorrect table position propagation during pitch/roll corrections. Affected users should contact for remedy.
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
RaySearch Laboratories is recalling RayStation/RayPlan 9B software version 9.1.0.933 because it may incorrectly propagate table positions when applying pitch or roll corrections. This could lead to inaccurate treatment planning in radiation therapy. The recall is for software used in medical radiation treatment planning systems.
The software may incorrectly calculate table positions when applying pitch or roll corrections, leading to inaccurate radiation treatment planning. Non-zero pitch or roll angles cause lateral, longitudinal, and vertical values to be incorrect, potentially affecting treatment precision.
Medical professionals using RayStation/RayPlan 9B software for radiation treatment planning. Patients receiving treatment with this software may be at risk if the software error affects treatment accuracy.
Check for RayStation/RayPlan 9B software version 9.1.0.933 with UDI 07350002010266. This software is used in medical radiation treatment planning systems.
Reach out to RaySearch Laboratories for guidance on the recall and remedy options.
The recall indicates a potential for inaccurate treatment planning, but the specific risk to patients is not detailed in the official notice.
Check for RayStation/RayPlan 9B software version 9.1.0.933 with UDI 07350002010266.
Contact RaySearch Laboratories for guidance on the recall and remedy options.
We’ve drafted a message you can send RaySearch Laboratories to request your refund or repair — edit it as you like.
Subject: Recall Z-1468-2020 — RaySearch Laboratories RaySearch Laboratories Hello, I own a RaySearch Laboratories that is covered by recall Z-1468-2020 from RaySearch Laboratories. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.