Respironics recalls Philips V60 Ventilators with Power Management PCBA part number 1055906 due to potential blower power loss causing ventilation issues. Affected units have specific serial numbers.
Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
Philips V60 Ventilators with Power Management PCBA part number 1055906 are being recalled because a solder connection failure may cause the ventilator to lose power or shut down unexpectedly, risking hypercarbia and hypoxemia. This could lead to serious breathing problems if the ventilator stops working during use.
A solder connection failure on the Power Management printed circuit board can cause the ventilator's blower to lose power, triggering alarms or shutting down unexpectedly. This may lead to hypercarbia (high carbon dioxide levels) and hypoxemia (low oxygen levels) if the ventilator stops working during critical use.
Patients using Philips V60 Ventilators with the specified Power Management PCBA part number are affected. Healthcare providers and caregivers responsible for these ventilators are most at risk of complications if the device fails.
Check if your Philips V60 Ventilator has Power Management PCBA part number 1055906. Affected units have serial numbers in specific ranges: 100002908 to 100027093, 100027096 to 100085389, 100085399 to 100109746, 100109868 to 201003390, or 201003392 to 201010952.
If your ventilator triggers a High Priority Check Vent alarm or shuts down unexpectedly, stop using it immediately.
Reach out to Respironics California, LLC for guidance on next steps if you own an affected ventilator.
Check for Power Management PCBA part number 1055906 on your Philips V60 Ventilator.
The ventilator may trigger a High Priority Check Vent alarm or shut down unexpectedly, potentially causing hypercarbia or hypoxemia.
Affected units have serial numbers in specific ranges: 100002908 to 100027093, 100027096 to 100085389, 100085399 to 100109746, 100109868 to 201003390, or 201003392 to 201010952.
We’ve drafted a message you can send Respironics to request your refund or repair — edit it as you like.
Subject: Recall Z-1622-2020 — Respironics Respironics Hello, I own a Respironics that is covered by recall Z-1622-2020 from Respironics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.