Maquet recalls Hemashield Platinum Woven Double Velour Vascular Grafts with potential small holes at seam lines causing intra-operative bleeding during surgery.
Potential small holes at the seam lines of Hemashield branches causing intra-operative bleeding at the seam lines of the branches of Hemashield grafts
Maquet is recalling Hemashield Platinum Vascular Grafts that may have small holes at seam lines, which could cause bleeding during surgery. This is a serious risk for patients undergoing vascular surgery.
Small holes at the seam lines of the grafts could cause bleeding during surgery. This may lead to significant blood loss and complications if the graft is used in procedures involving arterial repair or replacement.
Surgical teams using the Hemashield Platinum Vascular Grafts for artery repair or replacement are affected. Patients undergoing vascular surgery with these grafts are at immediate risk.
Check for the product code M00202175824P0 on the graft. The serial numbers listed in the recall include specific numbers like 1069833603 and 1073244538.
Immediately stop using the Hemashield Platinum Vascular Grafts with the listed serial numbers as they may cause bleeding during surgery.
The product code is M00202175824P0.
The affected serial numbers include 1069833603, 1069800979, 1069830915, and others listed in the recall notice.
The recall is due to potential small holes at the seam lines of the grafts that could cause intra-operative bleeding during surgery.
We’ve drafted a message you can send Maquet to request your refund or repair — edit it as you like.
Subject: Recall Z-1870-2020 — Maquet Maquet Hello, I own a Maquet that is covered by recall Z-1870-2020 from Maquet. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.