Siemens recalls Cios Spin VA30 mobile X-Ray systems with serial numbers 50030-50196. The main cable can disconnect during use, causing electrical shock if the X10 connector pins are touched. Affected units must be checked for error code 16305/80.
Main cable can be plugged or unplugged from the X10 connector which is part of the main unit (c-arm) at any time during clinical session. If the system displays the error message Err 16305 / 80, the unplugging of the main cable can cause an unsafe electrical voltage (125 V DC) at the X10 connector and result in an electrical shock if the user or other persons touch the contact pins of the X10 connector. These contact pins are accessible when the main cable is unplugged from the X10 connector
Siemens is recalling mobile X-ray systems that can disconnect during use, potentially causing electrical shock. If the system shows error code 16305/80, unplugging the main cable creates a dangerous voltage at the connector that could shock users.
The main cable can accidentally disconnect during X-ray procedures, creating a 125 V DC electrical shock hazard at the X10 connector. If the user touches the exposed contact pins after unplugging, they could receive an electrical shock. The system displays error code 16305/80 when this issue occurs.
Healthcare professionals using Siemens Cios Spin VA30 mobile X-ray systems with serial numbers starting from 50030 to 50196 are most at risk. Patients and staff near the system during clinical sessions could also be affected.
Check your Siemens Cios Spin VA30 mobile X-ray system for serial numbers listed: 50030, 50033, 50048, 50050, 50051, 50073, 50075, 50081, 50088, 50098, 50107, 50109, 50111, 50120, 50122, 50125, 50128, 50130, 50132, 50134, 50135, 50136, 50155, 50160, 50161, 50166, 50171, 50180, 50185, 50192, 50196. The system may show error code 16305/80 when the cable disconnects.
Look for error message 16305/80 on the system display, which indicates the main cable may have disconnected.
Immediately stop using the system if it displays error code 16305/80 to prevent electrical shock.
The system displays error code 16305/80 when the main cable disconnects during use.
Check your Siemens Cios Spin VA30 mobile X-ray system's serial number against the list: 50030 to 50196.
Immediately stop using the system and do not touch the X10 connector pins to avoid electrical shock.
We’ve drafted a message you can send Siemens to request your refund or repair — edit it as you like.
Subject: Recall Z-1959-2020 — Siemens Siemens Hello, I own a Siemens that is covered by recall Z-1959-2020 from Siemens. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.