Cardinal Health recalls NPWT Canister with Gel 300cc (model 47-4000) due to potential oversized O-ring causing installation failure. Prescription-only device for wound care.
Canisters potentially contain an oversized O-ring that cannot be installed into the mating connector on the device.
Cardinal Health is recalling NPWT Canister with Gel 300cc devices that may have an oversized O-ring. This could prevent proper installation, risking device failure during wound care treatment.
The canisters may have an oversized O-ring that won't fit correctly into the device connector. This could cause the system to fail during use, potentially leading to improper wound care.
Patients using prescription-only NPWT systems for wound care, especially those with chronic wounds, burns, or ulcers. Healthcare providers and home care users are most at risk.
Check for model number 47-4000 on the canister. Lot numbers 201921913, 201921915, 201921919, 201922009, 201922017, 201922019, 201922021, 201922118, 201922419, 201922120, 201922121, 201922421 are affected.
Verify the model number is 47-4000 and the lot number is listed in the affected numbers.
Reach out to Cardinal Health for specific guidance on whether your device needs repair or replacement.
The canisters may have an oversized O-ring that cannot be installed into the device connector, risking system failure during wound care.
No, but you should check if your canister is affected and contact Cardinal Health for next steps.
Contact Cardinal Health to determine if your device needs repair or replacement.
We’ve drafted a message you can send Cardinal Health to request your refund or repair — edit it as you like.
Subject: Recall Z-2543-2020 — Cardinal Health Cardinal Health Hello, I own a Cardinal Health that is covered by recall Z-2543-2020 from Cardinal Health. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.