Medtronic recalls Accurian RF Ablation cannulas due to probe insertion difficulty caused by catching on cannula hub lip. Affected units have specific product numbers and lot codes.
Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.
Medtronic is recalling Accurian RF Ablation cannulas because the probes can catch on a lip inside the cannula hub during insertion. This could lead to improper use or injury during medical procedures.
The probe's design causes it to catch on a lip inside the cannula hub during insertion. This could lead to improper placement or injury during medical procedures. The issue occurs with specific product numbers and lot codes.
Medical professionals using the Accurian RF Ablation system for nerve procedures are most at risk. Patients using this system are not directly affected.
Check for product number AC0007 and GTIN/UPN 20763000153568. Affected units include lot numbers BY19K031 and BY19L011.
Verify the product number AC0007 and GTIN/UPN 20763000153568 on the cannula packaging.
The probes can catch on a lip inside the cannula hub during insertion, which may cause improper placement or injury during medical procedures.
Check the product number and lot code on the cannula packaging. If it matches AC0007 with lot numbers BY19K031 or BY19L011, contact Medtronic for further instructions.
The recall indicates a potential issue with insertion difficulty, but the product is generally safe for use with proper handling. Affected units should be checked.
We’ve drafted a message you can send Medtronic to request your refund or repair — edit it as you like.
Subject: Recall Z-2691-2020 — Medtronic Medtronic Hello, I own a Medtronic that is covered by recall Z-2691-2020 from Medtronic. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.