CME America recalls CMExpress Microbore Set (REF: A120-161CYF) due to potential delivery inaccuracies in infusion therapy, risking over or under-dosing of medication.
Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy
CME America is recalling its CMExpress Microbore Set (REF: A120-161CYF) because it may deliver medication up to 17.5% over or 42.8% under the intended dose. This could lead to serious health risks if used for critical treatments.
The infusion sets may have delivery inaccuracies of up to 17.5% over or 42.8% under the intended dose. This can cause over-infusion or under-infusion of therapy, potentially leading to serious health consequences.
Patients using the CMExpress Microbore Set for infusion therapy are most at risk. The recall applies to all units within the product's expiry date.
Check for the product name 'CMExpress Microbore Set' and reference number A120-161CYF. The recall covers all units within the product's expiry date.
Identify the reference number A120-161CYF and expiry date on your infusion set.
Yes, the recall states the sets may cause over or under-dosing of medication, which could lead to serious health risks.
Look for the reference number A120-161CYF on the product label and check the expiry date.
The recall does not specify a remedy, so contact CME America for guidance on next steps.
We’ve drafted a message you can send CME America to request your refund or repair — edit it as you like.
Subject: Recall Z-2569-2020 — CME America CME America Hello, I own a CME America that is covered by recall Z-2569-2020 from CME America. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.