Siemens Healthcare recalls DCA Vantage handheld barcode scanners with specific model numbers due to potential barcode data errors when configured for Code 39 with check digit.
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.
Siemens Healthcare is recalling DCA Vantage handheld barcode scanners with model numbers DS430: HC0015BZZWW. The scanners may send incorrect barcode data if configured to run Code 39 with check digit, which could lead to errors in medical testing.
The scanner may send incorrect barcode data if configured to run Code 39 with check digit. This can cause the last character to be removed when scanning barcodes without a check digit, potentially leading to errors in medical testing.
Healthcare facilities using DCA Vantage analyzers with these scanners may be affected. Staff using the scanners for medical barcode scanning are most at risk of data errors.
Check for model numbers DS4308 with external accessory HC0015BZZWW on DCA Vantage Handheld Barcode Scanners. These scanners were sold in the US with specific serial numbers.
Verify if your DCA Vantage Analyzer is configured to run Code 39 with check digit.
The scanner may send incorrect barcode data if configured to run Code 39 with check digit, which could lead to errors in medical testing.
Model DS4308 with external accessory HC0015BZZWW.
No, but you should check if your scanner is configured for Code 39 with check digit to avoid data errors.
We’ve drafted a message you can send Siemens Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-2747-2020 — Siemens Healthcare Siemens Healthcare Hello, I own a Siemens Healthcare that is covered by recall Z-2747-2020 from Siemens Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.