ARJOHUNTLEIGH POLSKA recalls bariatric bed frames with modular junction box cables that may melt, affecting functionality. Serial numbers P0597830 to P0610470 are affected.
Modular Junction Box (MJB) cables can melt and affect the bed functionality.
ARJOHUNTLEIGH POLSKA is recalling bariatric bed frames that have modular junction box cables which can melt and cause malfunction. This recall affects beds used in acute and post-acute care settings, not home care.
The modular junction box cables in these beds can melt, which may cause the bed to malfunction. This is a risk in clinical environments where the bed is used for patient care, but not in home settings.
Healthcare facilities using the Citadel Plus Bariatric Bed Frame System in acute or post-acute care settings are most at risk. Home users are not intended to use this product.
Check for serial numbers between P0597830 and P0610470. Model numbers include FXX21A4A4AKFBB and others listed in the recall notice.
Look for serial numbers in the range P0597830 to P0610470 on the bed frame.
No, this bed is intended only for acute and post-acute care environments, not home care.
Affected models include FXX21A4A4AKFBB, FXX21A4D3AMFBA, and others listed in the recall notice.
If your serial number is outside the range P0597830 to P0610470, this recall does not apply to your bed.
We’ve drafted a message you can send ARJOHUNTLEIGH POLSKA to request your refund or repair — edit it as you like.
Subject: Recall Z-2821-2020 — ARJOHUNTLEIGH POLSKA ARJOHUNTLEIGH POLSKA Hello, I own a ARJOHUNTLEIGH POLSKA that is covered by recall Z-2821-2020 from ARJOHUNTLEIGH POLSKA. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.