Shanghai United Imaging Healthcare recalls uCT 760 CT scanners due to potential incorrect radiation exposure during multi-phase scans. Affected units have serial numbers 600229, 600230, 600232, 600271, 600282.
A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.
Shanghai United Imaging Healthcare is recalling uCT 760 CT scanners because they may display incorrect radiation exposure times during multi-phase scans. This could lead to unnecessary radiation exposure for medical staff performing scans.
The CT scanner's digital display panel may show incorrect radiation exposure times during multi-phase contrast scans. This could cause medical staff to receive unnecessary radiation exposure while performing scans.
Medical facilities using the uCT 760 CT scanners with serial numbers 600229, 600230, 600232, 600271, or 600282 are most at risk. Healthcare workers operating these scanners may receive unnecessary radiation exposure.
Check if your CT scanner has the serial numbers 600229, 600230, 600232, 600271, or 600282. The product is a CT scanner used for cross-sectional body imaging.
Verify your scanner's serial number against the list: 600229, 600230, 600232, 600271, 600282.
The CT scanner may display incorrect radiation exposure times during multi-phase scans, potentially causing unnecessary radiation exposure for medical staff.
Check your scanner's serial number against the list: 600229, 600230, 600232, 600271, 600282. If it matches, contact the manufacturer for further instructions.
The recall addresses potential unnecessary radiation exposure for medical staff, but the record does not specify a risk of serious injury.
We’ve drafted a message you can send Shanghai United Imaging Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0447-2021 — Shanghai United Imaging Healthcare Shanghai United Imaging Healthcare Hello, I own a Shanghai United Imaging Healthcare that is covered by recall Z-0447-2021 from Shanghai United Imaging Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.