Ameditech recalls 14-drug abuse test cups due to inconsistent drug detection results. Products may fail to detect calibrators and compounds as labeled, risking inaccurate drug screening.
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Ameditech recalls 14-drug abuse test cups that may not consistently detect drugs due to enzyme issues. This could lead to inaccurate drug screening results, which might affect proper testing outcomes.
The test cups use enzymes that sometimes fail to work properly. This means the cups might not detect all drugs or substances as labeled, leading to inconsistent results.
People using the Consult Diagnostics 14-drug abuse test cups for drug screening are affected. Those testing for drugs in themselves or others could be at risk of inaccurate results.
Look for the Consult Diagnostics 14-drug abuse test cups with model number 164-DXA14. The product includes all lots within expiry dates.
Verify the model number 164-DXA14 on the test cup packaging.
The test cups may not consistently detect all drugs due to enzyme failures, which could lead to inaccurate results.
The recall does not specify a remedy, so check the product details and contact Ameditech for further guidance.
The recall indicates potential inaccurate results, not direct safety risks like injury or fire.
We’ve drafted a message you can send Ameditech to request your refund or repair — edit it as you like.
Subject: Recall Z-3033-2020 — Ameditech Ameditech Hello, I own a Ameditech that is covered by recall Z-3033-2020 from Ameditech. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.