ESAOTE recalls AC/DC power supplies for specific ultrasound systems due to connector heating, smoke, and melting risks. Affected models include MyLabAlpha, MyLabGamma, and others with serial numbers below K17350001.
The AC/DC adapter (power supply) of the portable ultrasound systems listed below that may result in the connector heating up, generating smoke and melting. This condition is caused by the mechanical stress of the connection between the AC/DC power adapter and the mains cable over the time of use.
ESAOTE is recalling AC/DC power supplies for certain portable ultrasound systems because the connector can overheat, generate smoke, and melt over time. This poses a fire risk if the power supply is used without proper maintenance.
The power adapter connector heats up from repeated mechanical stress during normal use, which can cause smoke and melting. This creates a fire risk if the adapter is not properly maintained or replaced.
Owners of ESAOTE portable ultrasound systems with the listed models and serial numbers below K17350001 are affected. Medical professionals using these systems are most at risk due to potential fire hazards.
Check if your ultrasound system model is listed (e.g., MyLabAlpha, MyLabGamma) and has a serial number below K17350001. The power supply part number is 151004301 with XP Power model AHM180PS19-XE1057.
Identify if your power supply has part number 151004301 or XP Power model AHM180PS19-XE1057.
Replace the power supply with a compatible model from ESAOTE to prevent overheating and fire risks.
MyLabAlpha (101740000), MyLabGamma (101741000), MyLabOmega (101740004), MyLabSigma (101741004), MyLabSigma elite (101741007), MyLabAlphaVET (101740001), MyLabGammaVET (101741001), MyLabDelta (100743000), MyLabDeltaVET (100743001), and MyLabOmegaVET (101740005), MyLabSigmaVET (101741005).
Check for serial numbers below K17350001 and part number 151004301 with XP Power model AHM180PS19-XE1057.
Replace the power supply with a compatible model from ESAOTE to prevent overheating and fire risks.
We’ve drafted a message you can send ESAOTE to request your refund or repair — edit it as you like.
Subject: Recall Z-3053-2020 — ESAOTE ESAOTE Hello, I own a ESAOTE that is covered by recall Z-3053-2020 from ESAOTE. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.