Pioneer Surgical Technology recalls Streamline MIS Navigation System Taps with catalog numbers 04-CANNDRILL-MDN through 04-CANNTAP-85-MDN for binding issues during spinal surgery.
The Streamline Navigation System taps have experienced binding with other devices.
Pioneer Surgical Technology recalls specific Streamline MIS Navigation System Taps used in spinal surgery for binding with other devices. This could cause inaccurate screw placement during surgery, potentially leading to complications.
The taps can bind with other devices during spinal surgery, which may cause inaccurate screw placement. This could lead to complications if the surgeon cannot precisely locate anatomical structures.
Surgeons and medical facilities using the recalled Streamline Navigation System Taps for spinal procedures are most at risk. The recall affects specific catalog numbers sold in certain lots.
Check for catalog numbers starting with 04-CANNTAP-40-MDN through 04-CANNTAP-85-MDN or 04-CANNTAP-55-MDN, 04-CANNTAP-65-MDN, 04-CANNTAP-75-MDN, 04-CANNTAP-85-MDN. The product was sold in lots 371266 through 377806.
Identify if your taps have catalog numbers starting with 04-CANNTAP-40-MDN through 04-CANNTAP-85-MDN or 04-CANNTAP-55-MDN, 04-CANNTAP-65-MDN, 04-CANNTAP-75-MDN, 04-CANNTAP-85-MDN.
The Streamline Navigation System taps have experienced binding with other devices during spinal surgery, which could cause inaccurate screw placement.
Catalog numbers 04-CANNDRILL-MDN, 04-CANNTAP-40-MDN, 04-CANNTAP-45-MDN, 04-CANNTAP-55-MDN, 04-CANNTAP-65-MDN, 04-CANNTAP-75-MDN, 04-CANNTAP-85-MDN, 04-CTAP-TT-65-MDN, 04-S-FINDER-MDN, 05-SCREWINS-CL-MDN, and 05-SIMPDRIVER-MDN.
No action is required for the recall as the remedy is not specified. However, check if your taps are affected before using them in surgery.
We’ve drafted a message you can send Pioneer Surgical Technology to request your refund or repair — edit it as you like.
Subject: Recall Z-0299-2021 — Pioneer Surgical Technology Pioneer Surgical Technology Hello, I own a Pioneer Surgical Technology that is covered by recall Z-0299-2021 from Pioneer Surgical Technology. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.