Exactech recalls 6.5mm right Equinoxe Platform Fracture Stems with incorrect laser marking that could lead to improper use. Affected units have serial numbers 4971292-4971303.
The device may be incorrectly laser marked as a right stem, when it should have been laser marked as a left stem.
Exactech is recalling 6.5mm right Equinoxe Platform Fracture Stems that may be incorrectly laser marked as right when they should be left. This could cause improper use during medical procedures.
The stem may be marked as a right stem when it should be a left stem. This could lead to improper use during spinal fusion procedures if the wrong stem is implanted.
Orthopedic surgeons and medical professionals who use these stems for spinal procedures. Patients receiving these implants are at risk if the wrong stem is used.
Check for the product name 'Equinoxe Platform Fracture Stem, 6.5mm, Right' with serial numbers 4971292-4971303. The device is intended for cemented use only.
Verify the serial number is within 4971292-4971303 to confirm if it's affected.
The device may be incorrectly laser marked as a right stem when it should have been marked as a left stem.
Incorrect marking could lead to improper use during spinal fusion procedures if the wrong stem is implanted.
Check the serial number on the product; affected units have serial numbers 4971292-4971303.
We’ve drafted a message you can send Exactech to request your refund or repair — edit it as you like.
Subject: Recall Z-0308-2021 — Exactech Exactech Hello, I own a Exactech that is covered by recall Z-0308-2021 from Exactech. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.