Ortho Clinical Diagnostics recalls VITROS XT ALB-TP Slides due to potential false protein results. Affected units include all current, future, and past lots of product code 6844293.
Eltrombopag Interference for VITROS Chemistry Products TP Slides and VITROS XT Chemistry Products ALB-TP Slides
Ortho Clinical Diagnostics is recalling VITROS XT Chemistry Products ALB-TP Slides because they may give false protein test results. This could lead to incorrect health assessments if used in medical testing.
The slides may produce false protein measurements when used with certain blood tests, leading to inaccurate health assessments. This interference occurs due to eltmobag in the test samples.
Medical labs using VITROS XT 3400 or 7600 systems with ALB-TP Slides (product code 6844293) are affected. Patients receiving tests using these slides may be at risk of incorrect results.
Check for product code 6844293 on VITROS XT ALB-TP Slides. All current, future, and past lots are affected. The slides are used in VITROS XT 3400 and 7600 chemistry systems.
Look for product code 6844293 on the slides to confirm if they are affected.
Yes, all current, future, and past lots of product code 6844293 are affected.
The slides may produce false protein results, leading to inaccurate health assessments.
No, but you should check the product code to confirm if they are affected before continuing use.
We’ve drafted a message you can send Ortho Clinical Diagnostics to request your refund or repair — edit it as you like.
Subject: Recall Z-0845-2021 — Ortho Clinical Diagnostics Ortho Clinical Diagnostics Hello, I own a Ortho Clinical Diagnostics that is covered by recall Z-0845-2021 from Ortho Clinical Diagnostics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.