Abbott recalls i-STAT EG7+ cartridges that may show higher than expected calcium results when stored at room temperature for more than 3 days (lots *20100-*20339) or 7 days (lots *20340+).
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
Abbott is recalling i-STAT EG7+ cartridges that may give higher than expected calcium test results if stored at room temperature for more than 3 days (for some lot numbers) or 7 days (for other lot numbers). This could lead to inaccurate test results if used in medical settings.
Cartridges stored at room temperature for longer than the specified time may show higher calcium results than actual levels. This could lead to incorrect medical decisions if the test is used for diagnosis or treatment.
Healthcare professionals using Abbott i-STAT EG7+ cartridges for blood tests may be affected. Those who use the cartridges for calcium testing are most at risk of inaccurate results.
Check the lot number on the cartridge packaging. Lot numbers starting with *20100 to *20339 (with * = A, K, L, M, N, W, Y) may need replacement after 3 days at room temperature. Lot numbers *20340 and above may need replacement after 7 days at room temperature.
Look for the lot number on the cartridge packaging, which starts with a * followed by numbers (e.g., *20100 to *20339 or *20340+).
If stored at room temperature for more than 3 days (for lot numbers *20100 to *20339) or 7 days (for lot numbers *20340 and above), replace the cartridge immediately.
The recall is happening because i-STAT EG7+ cartridges may show higher than expected calcium test results when stored at room temperature for longer than the specified time.
Cartridges with lot numbers *20100 to *20339 can be stored at room temperature for up to 3 days. Lot numbers *20340 and above can be stored for up to 7 days.
If you stored the cartridges at room temperature for more than the specified time (3 days for some lots, 7 days for others), you should replace them immediately.
We’ve drafted a message you can send Abbott to request your refund or repair — edit it as you like.
Subject: Recall Z-0881-2021 — Abbott Abbott Hello, I own a Abbott that is covered by recall Z-0881-2021 from Abbott. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.