Olympus recalls MAJ-244 single-use replacement baskets due to potential inadequate packaging seals that could lead to infections if non-sterile equipment is used.
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Olympus is recalling single-use replacement baskets for their BML-3Q mechanical lithotripter. The recall is because a small percentage of these baskets have packaging seals that may not be fully sealed, increasing the risk of infection if used improperly.
The baskets have a small percentage with inadequate package seals due to partial heat sealing. This could allow contaminants to enter the sterile packaging, potentially causing infections if the baskets are used without proper sterility.
Healthcare professionals using the MAJ-244 replacement baskets for lithotripter procedures are most at risk. Patients using these baskets could face infections if non-sterile equipment is used.
Look for the MAJ-244 model number on the basket. These baskets are used with BML-3Q mechanical lithotripters and were sold with specific model numbers including 01K, 02K, 04K, 05K, 06K, 07K, 09K, 92K, 93K, 94K, 95K, 97K, and 9XK.
Inspect the packaging seal for any signs of partial heat sealing or damage.
The recall is because a small percentage of these baskets have inadequate package seals due to partial heat sealing, which could lead to infections if non-sterile equipment is used.
Look for the MAJ-244 model number on the basket, which is used with BML-3Q mechanical lithotripters and was sold with specific model numbers including 01K, 02K, 04K, 05K, 06K, 07K, 09K, 92K, 93K, 94K, 95K, 97K, and 9XK.
The recall record does not specify a remedy, so you should check the packaging seal for any signs of partial heat sealing or damage.
We’ve drafted a message you can send Olympus to request your refund or repair — edit it as you like.
Subject: Recall Z-1067-2021 — Olympus Olympus Hello, I own a Olympus that is covered by recall Z-1067-2021 from Olympus. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.