Infusion Pump Repair Model 8100 pumps serviced with non-original bezel repair parts between Jan 2021 and March 2021 may crack, causing infusion interruptions. Affected units have specific serial numbers.
Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Infusion Pump Repair's Model 8100 infusion pumps that were serviced with non-original bezel repair parts between January 2021 and March 2021 are recalled. Cracking of these repair parts could cause infusion interruptions, potentially leading to over or under infusion of medication.
Bezel repair parts installed by non-original manufacturers can crack over time, causing posts to separate from the bezel. This separation may lead to free flow, over infusion, under infusion, or interruption of medication delivery.
Users of Infusion Pump Repair Model 8100 pumps serviced between January 2021 and March 2021 with non-original bezel repair parts are affected. Patients receiving medical treatments through these pumps are most at risk.
Check if your pump was serviced by Infusion Pump Repair between January 15, 2021 and March 14, 2021. Look for specific serial numbers listed in the recall notice.
Compare your pump's serial number with the list of affected units provided in the recall notice.
Reach out to Infusion Pump Repair to confirm if your pump is affected and get instructions for next steps.
Bezel repair parts used by non-original manufacturers can crack over time, causing posts to separate from the bezel and potentially interrupting medication delivery.
Contact Infusion Pump Repair to confirm if your pump is affected and get instructions for next steps.
These pumps were distributed between January 15, 2021 and March 14, 2021.
We’ve drafted a message you can send Infusion Pump Repair to request your refund or repair — edit it as you like.
Subject: Recall Z-1544-2021 — Infusion Pump Repair Infusion Pump Repair Hello, I own a Infusion Pump Repair that is covered by recall Z-1544-2021 from Infusion Pump Repair. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.