Arcoma recalls Omnera 400A Digital Radiographic Systems from 2001-2003, 2006-2044, etc. due to liquid ingress causing short circuits that may move the overhead crane. Affected units have specific UDI/DI numbers.
There is the potential for the X-Ray system to short circuit due to an ingress of liquid in the display handles which may result in uncontrolled movement of the Overhead Tube Crane (OTC).
Arcoma is recalling Omnera 400A Digital Radiographic Systems that could short circuit from liquid entering the display handles. This might cause the overhead tube crane to move unexpectedly, posing a risk of injury.
Liquid can enter the display handles, causing a short circuit. This may lead to uncontrolled movement of the overhead tube crane, potentially causing injury to people nearby.
Medical facilities using Omnera 400A systems from 2001-2003, 2006-2044, and other years listed in the recall. Staff handling the equipment are most at risk if the short circuit occurs.
Check for the UDI/DI number 07350008750012 on the system. The product was sold between 2001-2003, 2006-2044, and other specific years listed in the recall.
Inspect the display handles for signs of liquid ingress or damage.
The Omnera 400A Digital Radiographic System with UDI/DI: 07350008750012.
From 2001-2003, 2006-2044, and other specific years listed in the recall.
Check for the UDI/DI number 07350008750012 on the system. The product was sold during the listed years.
We’ve drafted a message you can send Arcoma to request your refund or repair — edit it as you like.
Subject: Recall Z-1332-2022 — Arcoma Arcoma Hello, I own a Arcoma that is covered by recall Z-1332-2022 from Arcoma. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.