Arthrex recalls specific suture anchors that may not be properly sterilized due to a packaging seal issue. Affected units include model AR-2323BCC with lots 12758314 and 13022357.
Not properly sterilized, because the outer pouch seal on the Tyvek header was sealed closely to the foil pouch.
Arthrex is recalling certain suture anchors that may not be properly sterilized because the packaging seal was too tight. This could lead to infection risks during surgery if the product is used without proper sterilization.
The outer pouch seal on the Tyvek header was sealed too closely to the foil pouch, which could prevent proper sterilization. If the product is used without adequate sterilization, it may cause infections during surgery.
Surgical teams and hospitals that use these specific suture anchors for procedures. Patients undergoing surgery with these anchors are at risk of infection if the product is not properly sterilized.
Check for the model number AR-2323BCC with size 5.5 mm x 19.1 mm, Closed Eyelet, and lot numbers 12758314 or 13022357. These suture anchors were sold by Arthrex.
Identify if your suture anchor has the model number AR-2323BCC with lot numbers 12758314 or 13022357.
The outer pouch seal on the Tyvek header was sealed too closely to the foil pouch, which could prevent proper sterilization.
Check for the model number AR-2323BCC with size 5.5 mm x 19.1 mm, Closed Eyelet, and lot numbers 12758314 or 13022357.
The recall record does not specify a remedy, so contact Arthrex directly for guidance.
We’ve drafted a message you can send Arthrex to request your refund or repair — edit it as you like.
Subject: Recall Z-2107-2021 — Arthrex Arthrex Hello, I own a Arthrex that is covered by recall Z-2107-2021 from Arthrex. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.