Fujifilm Irvine Scientific recalls Vit Kit-Freeze due to incorrect expiration date labeling on components. This cryopreservation kit for oocytes and embryos may lead to improper use if expired dates are misread.
Due to a component in kit being labeled with the incorrect Expiration Date
Fujifilm Irvine Scientific is recalling Vit Kit-Freeze, a cryopreservation kit used for freezing human oocytes and embryos. The recall is due to an incorrect expiration date label on one of the components, which could lead to improper use if the expiration date is misread.
The kit contains a component labeled with the wrong expiration date. If the expiration date is misread, the kit may be used after it has expired, potentially leading to improper freezing of biological samples.
Individuals using Vit Kit-Freeze for assisted reproductive procedures, such as oocyte or embryo freezing, are likely affected. Those who have purchased the kit from the manufacturer or authorized sellers are at risk.
Check for the product name 'Vit Kit-Freeze' and catalog numbers 90133-DCOS or 90133-DSOC. The kit includes 9x1 mL of Equilibration Solution-ES and Vitrification Solution-VS.
Verify the expiration date on the component labeled with the incorrect date to ensure it is valid.
Vit Kit-Freeze is intended for use in assisted reproductive procedures for vitrification and storage of human oocytes, pronuclear zygotes, and embryos up to the blastocyst stage.
The recall is due to an incorrect expiration date label, which could lead to improper use if the expiration date is misread. There is no indication of serious injury risk in normal use.
Look for the product name 'Vit Kit-Freeze' with catalog numbers 90133-DCOS or 90133-DSOC. The kit includes 9x1 mL of Equilibration Solution-ES and Vitrification Solution-VS.
We’ve drafted a message you can send Fujifilm Irvine Scientific to request your refund or repair — edit it as you like.
Subject: Recall Z-0248-2022 — Fujifilm Irvine Scientific Fujifilm Irvine Scientific Hello, I own a Fujifilm Irvine Scientific that is covered by recall Z-0248-2022 from Fujifilm Irvine Scientific. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.