Innovative Orthopedic Technologies recalls ARCH Leg Positioning System model 6813010 with serial numbers 6813010.0066-0076 due to locking pins failing to engage under load, potentially causing injury during use.
Due to the holes not being bored to the specific diameter, the locking pins are unable to engage under a load.
Innovative Orthopedic Technologies is recalling their ARCH Leg Positioning System model 6813010 with specific serial numbers because the locking pins may not engage properly under load. This could lead to the device failing during use, potentially causing injury.
The locking pins in the device may not engage properly under load because the holes were not bored to the correct diameter. This could cause the device to fail during use, potentially leading to injury.
Patients using the ARCH Leg Positioning System for medical purposes are affected. Those with the specific model and serial numbers listed are at risk of injury if the device fails during use.
Check for the model number 6813010 and serial numbers 6813010.0066 through 6813010.0076 on the device. The product was sold by Innovative Orthopedic Technologies.
Look for the model number 6813010 and serial numbers 6813010.0066 through 6813010.0076 on the device.
The affected model is 6813010 with serial numbers 6813010.0066, 6813010.0067, 6813010.0069, 6813010.0070, 6813010.0071, 6813010.0072, 6813010.0073, 6813010.0074, 6813010.0075, and 6813010.0076.
The locking pins may not engage properly under load because the holes were not bored to the specific diameter, which could cause the device to fail during use.
Check the device for the model number 6813010 and serial numbers 6813010.0066 through 6813010.0076. If you have the affected model, contact Innovative Orthopedic Technologies for further instructions.
We’ve drafted a message you can send Innovative Orthopedic Technologies to request your refund or repair — edit it as you like.
Subject: Recall Z-0072-2022 — Innovative Orthopedic Technologies Innovative Orthopedic Technologies Hello, I own a Innovative Orthopedic Technologies that is covered by recall Z-0072-2022 from Innovative Orthopedic Technologies. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.