Philips Respironics recalls V60 ventilators with expired adhesive that could cause silent shutdowns or alarm failures. Affected units have specific serial numbers and part numbers.
A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).
Philips Respironics is recalling a limited number of V60 ventilators due to expired adhesive in the bracket assembly. If the adhesive fails, the ventilator may stop working without warning, risking patient safety.
The expired adhesive can cause the bracket to become loose, potentially damaging capacitors. This may lead to the ventilator stopping ventilation without triggering alarms (silent shutdown) or activating both visual and audible alarms.
Patients using the recalled ventilators are most at risk. The affected units are limited to specific serial numbers and part numbers.
Check for the part number 1053617 on the ventilator. Affected units have specific serial numbers listed in the recall notice, including MX0016593 and other codes.
Look for the serial numbers listed in the recall: MX0016593, MX0016797, MX0016824, MX0016827, MX0016830, MX0016839, MX0016840, MX0016841, MX0016845, MX0016853, MX0016867, MX0016870, MX0016871, MX0016872, MX0016875, MX0016878.
The ventilator was assembled with expired adhesive, which could cause the bracket to become loose and damage capacitors, potentially leading to silent shutdowns or alarm failures.
Check the serial numbers listed in the recall notice to see if your unit is affected. If it is, contact Philips Respironics for further instructions.
The recall states that the ventilator may stop working without warning, which could risk patient safety, but the remedy is not specified in the record.
We’ve drafted a message you can send Philips Respironics to request your refund or repair — edit it as you like.
Subject: Recall Z-0662-2022 — Philips Respironics Philips Respironics Hello, I own a Philips Respironics that is covered by recall Z-0662-2022 from Philips Respironics. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.