B Braun recalls Conductivity Sensor (BIC) Version 2 for Dialog and Dialog+ hemodialysis systems due to potential cracks causing leaks. This may lead to improper ultrafiltration in dialysis treatments.
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
B Braun is recalling Conductivity Sensor (BIC) Version 2 used with Dialog and Dialog+ hemodialysis machines because cracks could cause leaks in the dialysis circuit. This might lead to either too much or too little ultrafiltration, which is critical for dialysis treatment safety.
The conductivity sensors may develop cracks during use, which can cause leaks in the dialysis circuit. If a leak occurs, it might lead to either excessive or insufficient ultrafiltration depending on whether the machine has a Dialysis Fluid filter.
Hemodialysis patients using Dialog or Dialog+ machines with the recalled Conductivity Sensor (BIC) Version 2 are most at risk. Patients using these machines with the specific model numbers listed in the recall are affected.
Check if your hemodialysis machine uses the Dialog or Dialog+ system and has the Conductivity Sensor (BIC) Version 2 with Article Code 3456102A. The affected models include 710200L, 710200S, 710200U, 710200K, 710500L, and 710500K.
Identify if your hemodialysis machine is one of the affected models listed in the recall: 710200L, 710200S, 710200U, 710200K, 710500L, or 710500K.
Reach out to B Braun Medical Inc for specific guidance on whether your Conductivity Sensor (BIC) Version 2 is affected and what steps to take next.
B Braun learned of potential cracks in the Conductivity Sensor (BIC) Version 2 that could cause leaks in the dialysis circuit, leading to improper ultrafiltration.
Cracks in the sensor may cause leaks in the dialysis circuit, which could result in either excessive or insufficient ultrafiltration depending on whether the machine has a Dialysis Fluid filter.
Check if your hemodialysis machine uses Dialog or Dialog+ systems and has the Conductivity Sensor (BIC) Version 2 with Article Code 3456102A. Affected models include 710200L, 710200S, 710200U, 710200K, 710500L, and 710500K.
We’ve drafted a message you can send B Braun to request your refund or repair — edit it as you like.
Subject: Recall Z-0872-2022 — B Braun B Braun Hello, I own a B Braun that is covered by recall Z-0872-2022 from B Braun. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.