Draeger recalls Globe-Trotter and Globe-Trotter IC neonatal transport systems with model numbers MS20508 and MS20509 due to an unrated oxygen supply component that could fail under pressure.
The firm has initiated this recall to replace component on devices distributed containing a component within the oxygen supply system that is not rated to the working pressure of the system.
Draeger is recalling specific neonatal transport systems (Globe-Trotter and Globe-Trotter IC models MS20508 and MS20509) because a component in the oxygen supply system isn't rated for the system's working pressure. This could lead to failure during use, potentially causing harm to infants.
The oxygen supply component in these systems isn't rated to handle the working pressure of the device. This means it could fail or leak oxygen under normal operating conditions, potentially causing harm to infants during transport.
Healthcare providers using Draeger Globe-Trotter or Globe-Trotter IC neonatal transport systems with model numbers MS20508 or MS20509 are affected. Infants receiving care with these systems are at risk if the oxygen supply component fails.
Check for model numbers MS20508 or MS20509 on the device. The product was sold in specific countries with serial numbers listed in the recall notice, including El Salvador, Mexico, Chile, Saudi Arabia, Kuwait, India, Taiwan, Germany, Qatar, Spain, Hungary, Belarus, United Kingdom, and Italy.
Identify the model number on your device to confirm if it is MS20508 or MS20509.
Reach out to Draeger Medical Systems, Inc. for instructions on how to replace the component or get a replacement device.
Draeger is recalling these devices because a component in the oxygen supply system isn't rated for the system's working pressure, which could lead to failure during use.
Model numbers MS20508 and MS20509 for the Globe-Trotter and Globe-Trotter IC neonatal transport systems.
The devices were sold in El Salvador, Mexico, Chile, Saudi Arabia, Kuwait, India, Taiwan, Germany, Qatar, Spain, Hungary, Belarus, United Kingdom, and Italy with specific serial numbers.
We’ve drafted a message you can send Draeger to request your refund or repair — edit it as you like.
Subject: Recall Z-1365-2022 — Draeger Draeger Hello, I own a Draeger that is covered by recall Z-1365-2022 from Draeger. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.