Recall of PICC line and central line packs due to unvalidated sterilization cycles. Affected items have specific tray and lot numbers. Stop using immediately to prevent infection risks.
Product was exposed to multiple sterilization cycles without validation for multiple exposures.
American Contract Systems is recalling certain PICC line and central line packs that were sterilized without proper validation. This could lead to infections if the medical devices are used without proper sterilization.
The packs were exposed to multiple sterilization cycles without proper validation. This means the medical devices may not be sterile, increasing the risk of infection when used in medical procedures.
Healthcare providers and patients using the recalled PICC line and central line packs. Those with the specific tray and lot numbers listed are most at risk.
Check the tray number, sterilization lot, lot number, and bag serial number on the pack. Affected items include those with tray numbers LSRPICCLINEH (2105051) or UICL44M (210:142).
Discard the recalled PICC line or central line packs if you have the specific tray and lot numbers listed.
The packs were sterilized without proper validation, which could lead to infections if used in medical procedures.
Check the tray number, sterilization lot, lot number, and bag serial number. Affected items include those with tray numbers LSRPICCLINEH (2105051) or UICL44M (2107142).
Discard the packs immediately to prevent infection risks.
We’ve drafted a message you can send American Contract Systems to request your refund or repair — edit it as you like.
Subject: Recall Z-1692-2022 — American Contract Systems American Contract Systems Hello, I own a American Contract Systems that is covered by recall Z-1692-2022 from American Contract Systems. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.