Baxter Healthcare recalls ABACUS TPN calculation software versions 3.1-3.3 due to potential medication errors. Affected users should contact Baxter for details.
There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
Baxter Healthcare is recalling specific versions of their ABACUS TPN (Total Parenteral Nutrition) calculation software because there's a risk of medication errors when used with certain medical devices. This could lead to incorrect dosing of critical medications for patients receiving intravenous nutrition.
The software may produce incorrect medication calculations when used with ExactaMix devices, potentially leading to improper dosing of critical nutrients and medications. This risk occurs during the software's normal operation when entering patient data and calculating medication requirements.
Healthcare providers using the recalled software with ExactaMix automated compounding devices (ExactaMix 1200 or 2400) are most at risk. Patients receiving intravenous nutrition through this system could be affected.
Check for product codes 8300-0167 (V3.1 CE), 8300-0168 (V3.1 SE), 8300-0169 (V3.1 ME), 8300-0191 (V3.2 CE), 8300-0192 (V3.2 SE), 8300-0193 (V3.2 ME), 8300-3391 (V3.3 CE), 8300-3392 (V3.3 SE), or 8300-3393 (V3.3 ME). The software is commonly used with ExactaMix 1200 or 2400 devices.
Verify your software version by checking the product code on the device or documentation.
Reach out to Baxter Healthcare for specific instructions on how to update or replace the software.
Product codes 8300-0167 (V3.1 CE), 8300-0168 (V3.1 SE), 8300-0169 (V3.1 ME), 8300-0191 (V3.2 CE), 8300-0192 (V3.2 SE), 8300-0193 (V3.2 ME), 8300-3391 (V3.3 CE), 8300-3392 (V3.3 SE), and 8300-3393 (V3.3 ME).
The software may produce incorrect medication calculations when used with ExactaMix devices, potentially leading to improper dosing of critical nutrients and medications.
No, but you should check your software version and contact Baxter Healthcare for instructions on updating or replacing it.
We’ve drafted a message you can send Baxter Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-1376-2022 — Baxter Healthcare Baxter Healthcare Hello, I own a Baxter Healthcare that is covered by recall Z-1376-2022 from Baxter Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.