Abbott recalls ARCHITECT STAT Tropon: Reagent Kit REF 2K41-27 with insufficient heterophilic blocking agents. Stop using if you have the specific lot number.
One lot of reagent was manufactured with insufficient quantities of the heterophilic blocking agents in the conjugate component.
Abbott is recalling a specific lot of their ARCHITECT STAT Troponin-I Reagent Kit due to insufficient heterophilic blocking agents in the reagent. This could lead to inaccurate test results if used without the fix.
The reagent kit contains insufficient heterophilic blocking agents, which are critical for accurate troponin-I testing. This could cause false negative results if the kit is used without the fix.
Lab technicians and healthcare providers who use the ARCHITECT STAT Troponin-I Reagent Kit REF 2K41-27. Those with the specific lot number 63718UN22 are most at risk.
Check the lot number on the reagent kit label: 63718UN22. The product is sold under the reference number 2K41-27 and was manufactured in lot 63718UN22.
Look for the lot number 63718UN22 on the reagent kit label.
No, the specific lot 63718UN22 has insufficient heterophilic blocking agents and should not be used without the fix.
Check the lot number; if it's 63718UN22, stop using it and contact Abbott for instructions.
The product has the lot number 63718UN22 and reference number 2K41-27.
We’ve drafted a message you can send Abbott to request your refund or repair — edit it as you like.
Subject: Recall Z-1660-2022 — Abbott Abbott Hello, I own a Abbott that is covered by recall Z-1660-2022 from Abbott. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.