MEDLINE recalls Clearpro Trach T-Piece Closed Suction Catheters with potential for connector to separate during patient suctioning, causing injury. Affected units have specific lot numbers and item references.
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
MEDLINE is recalling Clearpro Trach T-Piece Closed Suction Catheters that may come apart during use, risking injury to patients. This recall affects specific lot numbers and model types sold in the U.S. The hazard occurs when the catheter is pulled out of the artificial airway.
The suction catheter's green connector near the thumb valve can separate from the catheter when pulled out of the patient's airway during suctioning. This separation could cause the catheter to become dislodged, potentially leading to injury or complications.
Healthcare professionals using the catheter for patient suctioning are most at risk. The recall applies to specific lot numbers and model types sold in the U.S.
Check for the item number/REF DYNCPTP14 or DYNCPTP14T on the catheter label. Affected units have specific lot numbers: 6922030011, 6921050011, 6921060011, 6921040012, 6921110011, 6921120011, 6921120012, 6922010011, 6922010012, 6922010031, 6922020031, 6922030011.
Look for item numbers DYNCPTP14 or DYNCPTP14T on the catheter label to confirm if it's affected.
If the product is affected, contact MEDLINE directly for a replacement or repair.
The green connector near the thumb valve can separate from the catheter when pulled out of the patient's airway during suctioning, potentially causing injury.
Affected lot numbers include 6922030011, 6921050011, 6921060011, 6921040012, 6921110011, 6921120011, 6921120012, 6922010011, 6922010012, 6922010031, 6922020031, 6922030011.
Check the label for item numbers DYNCPTP14 or DYNCPTP14T and verify the lot numbers listed in the recall notice.
We’ve drafted a message you can send MEDLINE to request your refund or repair — edit it as you like.
Subject: Recall Z-1857-2022 — MEDLINE MEDLINE Hello, I own a MEDLINE that is covered by recall Z-1857-2022 from MEDLINE. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.