Hologic recalls BioZorb LP markers with expired labels that show incorrect dates. The product may be used incorrectly due to outdated expiration dates on removable labels.
Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is actually manufacturing date, and it's different from the one on the shelf box label, which is the correct expiration date.
Hologic recalls BioZorb LP markers that have removable labels showing an expired date (January 20, 2022) instead of the correct expiration date (July 8, 2024). Using these markers could lead to improper medical procedures if the expiration date is mistaken for the actual expiration.
The removable label on the BioZorb LP marker shows an expiration date of January 20, 2022, but the correct expiration date is July 8, 2024. Using markers with this incorrect label could lead to improper medical procedures if the expiration date is mistaken for the actual expiration.
Medical professionals who use the BioZorb LP markers for soft tissue visualization may be affected. Patients receiving procedures with these markers could be at risk if the incorrect expiration date leads to improper use.
Check the removable label on the BioZorb LP marker for an expiration date of January 20, 2022. The correct expiration date is July 8, 2024, as listed on the shelf box label.
Verify the expiration date on the removable label matches the correct expiration date of July 8, 2024 as shown on the shelf box label.
The removable label shows an expiration date of January 20, 2022, but the correct expiration date is July 8, 2024. This is due to a manufacturing error where the label's expiration date was set to the manufacturing date.
Check the removable label for an expiration date of January 20, 2022. If it matches, contact Hologic for guidance on next steps.
The hazard is low risk as the issue involves incorrect labeling rather than a physical defect that could cause immediate harm.
We’ve drafted a message you can send Hologic to request your refund or repair — edit it as you like.
Subject: Recall Z-1858-2022 — Hologic Hologic Hello, I own a Hologic that is covered by recall Z-1858-2022 from Hologic. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.