BD Biosciences recalls BD Trucount Tubes (Cat. No. 340334) due to partial label detachment causing tubes to stick together or automation errors. Affected units include specific lot numbers.
Due to partial label detachment on tubes. This can cause tubes to become stuck together in the pouch or cause automation errors when used with the automated loader for the wash assistant. Firm root cause investigation confirmed that label detachment is due to an adhesive change by a tier 2 supplier.
BD Biosciences is recalling BD Trucount Tubes (Cat. No. 340334) because partial label detachment can cause tubes to become stuck together or lead to automation errors when used with certain lab equipment. This recall affects specific lot numbers produced between 2031144 and 2237016.
Partial label detachment on the tubes can cause them to stick together in the pouch or result in automation errors when used with the BD FACS Loader. This issue is due to an adhesive change by a tier 2 supplier.
Lab users who own or use BD Trucount Tubes (Cat. No. 340334) with specific lot numbers are most at risk. The recall primarily impacts laboratory settings using these tubes for blood cell counting.
Check for BD Trucount Tubes with catalog number 340334 and lot numbers listed in the recall (2031144 to 2237016). These tubes are used for blood cell counting in laboratory settings.
Identify the lot number on the tube to see if it is listed in the recall (2031144 to 2237016).
Reach out to BD Biosciences for further assistance if you have the affected lot numbers.
Partial label detachment on the tubes can cause them to stick together in the pouch or lead to automation errors when used with the BD FACS Loader. This is due to an adhesive change by a tier 2 supplier.
Check the lot number on the tube to see if it is listed in the recall (2031144 to 2237016). Contact BD Biosciences for further assistance.
The recall indicates a risk of tubes sticking together or automation errors, but no immediate safety hazards like injury or fire are mentioned.
We’ve drafted a message you can send BD Biosciences to request your refund or repair — edit it as you like.
Subject: Recall Z-0171-2023 — BD Biosciences BD Biosciences Hello, I own a BD Biosciences that is covered by recall Z-0171-2023 from BD Biosciences. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.