Ferno recalls Power X1 Ambulance Cots with 36V Li-Ion Batteries manufactured through 2021, week 15. Batteries may become unstable if not maintained or subjected to drops, power washing, or abuse, risking fire or injury.
Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing directly into battery contacts or abuse, may result in a fire and/or injury
Ferno is recalling Power X1 Ambulance Cots with 36V Li-Ion Batteries made through 2021, week 15. These batteries can become unstable if not properly maintained, dropped, or exposed to power washing, potentially causing fires or injuries.
The lithium-ion batteries can become unstable if not properly maintained, subjected to repeated drops, power washing directly into battery contacts, or abuse. This instability may lead to fires or injuries.
Users of the Power X1 Ambulance Cot with 36V Li-Ion Batteries manufactured through 2021, week 15 are affected. Those who use the cot for medical purposes or in environments where the battery might be dropped or exposed to water are at higher risk.
Check if your Power X1 Ambulance Cot has a serial number starting with '18F003862' through '21S024868' (all numbers made through 2021, week 15). The battery model is #0822484/0822483.
Look for the battery serial number on the cot's label; it should start with '00190790005474' for the 36V Li-Ion battery.
If the battery has been dropped, power washed, or abused, stop using the cot immediately to prevent fire or injury.
The recall number is not specified in the official notice.
Check the serial number on your cot; affected units have serial numbers starting with '18F003862' through '21S024868' (all made through 2021, week 15).
If your cot has a battery that was dropped, power washed, or abused, stop using it immediately. Otherwise, contact Ferno for further instructions.
We’ve drafted a message you can send Ferno to request your refund or repair — edit it as you like.
Subject: Recall Z-0184-2023 — Ferno Ferno Hello, I own a Ferno that is covered by recall Z-0184-2023 from Ferno. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.