Carefusion recalls bipolar forceps with missing instructions for power supply and cleaning. Owners should check model numbers and contact Carefusion for details.
BD has determined that the Instructions For Use (IFU) provided by BD is missing certain content that is contained in the manufacturer's IFU. This missing content is related to the interface with the device power supply, as well as cleaning and maintenance instructions. Refer to the attached V. Mueller" Bipolar Forcep IFU.
Carefusion is recalling bipolar forceps that lack critical instructions for power supply and cleaning. This could lead to improper use or damage if the device isn't maintained correctly.
The missing instructions could cause improper power supply connections or inadequate cleaning, potentially leading to device malfunction or infection risks.
Medical professionals using these forceps for procedures are most at risk. The recall affects specific models sold with certain lot numbers.
Check for model number 10885403040719 on the product. The forceps have a 1.5mm tip and 7-3/4" (195mm) length. All lots of this specific model are affected.
Look for the model number 10885403040719 on the forceps to confirm if it's recalled.
The instructions for use are missing content related to power supply and cleaning, which could lead to improper use or damage.
No, but you should check the model number and contact Carefusion for further instructions.
Check for model number 10885403040719 on the product. All lots of this specific model are affected.
We’ve drafted a message you can send Carefusion to request your refund or repair — edit it as you like.
Subject: Recall Z-0374-2023 — Carefusion Carefusion Hello, I own a Carefusion that is covered by recall Z-0374-2023 from Carefusion. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.