Biomerieux recalls ETEST Minocycline MC 256 US S30 (Catalog 412407) due to potential performance issues from temperature and time exposure. Affected units include specific batch numbers.
As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Biomerieux is recalling ETEST Clinical Minocycline MC 256 US S30 products because they may not work properly if stored outside recommended temperature and time ranges. This could lead to inaccurate test results if used incorrectly.
The product may not perform correctly if stored outside the manufacturer's specified temperature and time ranges. This could result in inaccurate test results if used without proper handling.
Healthcare professionals and labs using the ETEST Minocycline MC 256 US S30 products are most likely affected. The risk is low as the issue relates to storage conditions, not immediate safety hazards.
Check for the product name 'ETEST Clinical Minocycline MC 256 US S30' and catalog number 412407. Affected units have batch numbers starting with 1009434000.
Verify the product name, catalog number, and batch numbers to confirm if your item is affected.
No, the recall is due to potential performance issues from improper storage, not immediate safety hazards.
Check the product details to see if it's affected; no action is needed if it's not in the recall list.
Look for the product name 'ETEST Clinical Minocycline MC 256 US S30' and catalog number 412407 with batch numbers starting with 1009434000.
We’ve drafted a message you can send Biomerieux to request your refund or repair — edit it as you like.
Subject: Recall Z-0828-2023 — Biomerieux Biomerieux Hello, I own a Biomerieux that is covered by recall Z-0828-2023 from Biomerieux. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.