Mobius Imaging recalls MobiCT-32 AIRO Mobile CT Scanners with specific serial numbers due to battery fire risk from bypassed charging systems. Affected units require battery replacement.
Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.
Mobius Imaging is recalling specific MobiCT-32 AIRO Mobile CT Scanners that have battery monitoring systems bypassed during charging. This could lead to battery fires if the batteries are not replaced as required.
Bypassing the battery monitoring system during charging can cause the batteries to overheat and potentially catch fire. This risk occurs when the battery is charged without the safety checks that normally prevent overheating.
Healthcare facilities using the MobiCT-32 AIRO Mobile CT Scanners with serial numbers AIRO-0124, AIRO-0145, or AIRO-0288 are most at risk. These units were manufactured with battery monitoring systems that could be bypassed during charging.
Check for the serial numbers AIRO-0124, AIRO-0145, or AIRO-0288 on your MobiCT-32 AIRO Mobile CT Scanner. The product was sold with software version 2.1.1 and has a specific UDI-DI code.
Verify if the battery monitoring system was bypassed by checking for signs of overheating or unusual behavior during charging.
Contact Mobius Imaging to arrange for battery replacement as the batteries require replacement due to the bypassed monitoring system.
The recall is due to a risk of battery fire if the battery monitoring system was bypassed during charging. Affected units require battery replacement.
Check for the serial numbers AIRO-0124, AIRO-0145, or AIRO-0288 on your MobiCT-32 AIRO Mobile CT Scanner. The product was sold with software version 2.1.1.
No, but you should check if the battery monitoring system was bypassed and replace the batteries if necessary. The recall does not require immediate shutdown.
We’ve drafted a message you can send Mobius Imaging to request your refund or repair — edit it as you like.
Subject: Recall Z-0344-2023 — Mobius Imaging Mobius Imaging Hello, I own a Mobius Imaging that is covered by recall Z-0344-2023 from Mobius Imaging. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.